System Setup
Instructions for Use Apollo SW 4.5n 87
Part Number: 9053586, 3rd edition
– Do not attempt to drag the device over hoses,
cables, or other obstructions on the floor.
To increase toppling stability:
– Remove all monitors and devices from the upper
storage area.
– Dismantle any additional mounted devices on
swivel arms or on the upper side of the device
(e.g. patient monitoring, data management
systems, syringe pumps, etc.)
– Remove vaporizers and gas cylinders.
– Clear the writing table and fold it down
completely.
– Position the breathing bag arm close to the
device.
– Push in the ventilator module and drawers.
WARNING!
Risk of injury
If handled incorrectly, the anesthesia machine
may become top-heavy and tip over causing
injury to the patient and/or user.
Observe the following points to prevent this
hazard.
CAUTION!
Risk of physical injury
To avoid physical injury, e.g. pinching, pay special
attention to edges, moving parts and corners when
working with
– drawers,
– the ventilator module,
– doors,
– the writing tray,
– swivel arms for mounted devices,
– gas cylinders,
– vaporizer units,
– CLIC absorbers and CLIC adapters,
as well as other accessories.