2. [Operating mode] - [Pulse]
- select.
3.
[Press the Next button] .
ð
Screen change for setting: ‘Pulse interval’.
4. Enter pulse interval (ml, l / pulse) of the flowmeter being used.
5. Select the field for entering the value after the decimal point.
ð
The colour changes to dark blue, which indicates that the field is ready for the
entry
.
6. Set the decimal point with the scrollbar
.
7.
Press the Next button.
ð
The screen changes to the ‘Pulse memory’ prompt.
8. Select whether the ‘pulse memory’
is to be used.
ð
If you select the ‘Pulse memory’, a
appears in the selection box.
9.
Press the Next
button.
ð
Screen change to the setting for ‘concentration’.
10. Enter solution concentration (% / ppm) of the flowmeter being used.
11. Select the field for entering the value after the decimal point.
ð
The colour changes to dark blue, which indicates that the field is ready for the
entry
.
12. Set the decimal point with the scrollbar
.
13.
Press the
button.
ð
Settings are saved, and the screen returns to the [Operating mode] .
14.
Press the button.
ð
The screen switches back to the overview [Main menu] .
15.
Press the button.
ð
The screen switches back to the [Operating display] .
16. Keep the displayed value (% / ppm) of the concentration pressed down for approx. 3
seconds.
ð
The configuration page for converting the unit (% / ppm) is displayed.
17. T
ap the required unit (% / ppm) of the concentration setting and select the required
unit.
18.
Press the
button.
ð
The screen switches back to the concentration setting.
19. If necessary
, adjust the concentration value with the newly set unit.
20.
Press the
button.
ð
The screen switches back to the [Operating display] and displays the newly set
unit.
Control / software
75 Rev. 2-05.2018