Troubleshooting
Table10-2Statusalarmsandrecommendedactions
continued
AlarmcodeDescriptionCauseRecommendedactions
A105
Slugow
Thedensityhas
exceededthe
user-denedslug
(density)limits.
•Checkforslugow.SeeSection10.24.
A106BurstmodeenabledThedeviceisinHART
burstmode.
•Noactionrequired.
•Ifdesired,youcanrecongurethealarm
severityleveltoIgnore.
A107PowerresetoccurredThetransmitterhas
beenrestarted.
•Noactionrequired.
•Ifdesired,youcanrecongurethealarm
severityleveltoIgnore.
A108Event1triggered
•Noactionrequired.
•Revieweventcongurationifyoubelieve
theeventwastriggerederroneously.
A109Event2triggered
•Noactionrequired.
•Revieweventcongurationifyoubelieve
theeventwastriggerederroneously.
A110Frequencyoutput
saturated
Thecalculated
frequencyoutputis
outsideofthelinear
range.
•Checkthefrequencyoutputscaling.See
Section10.20.
•Checkprocessconditions.Actual
conditionsmaybeoutsideofthenormally
expectedconditionsforwhichtheoutput
iscongured.
•Verifyprocessconditions,checking
especiallyforairintheowtubes,tubes
notlled,foreignmaterialinthetubes,or
coatinginthetubes.
•Verifythatthemeasurementunitsare
conguredcorrectlyforyourapplication.
•Purgetheowtubes.
A111
FrequencyoutputxedUserhasxedthe
frequencyoutput.
•Checkthatthetransmitterisinlooptest
mode.
•Checkthattheoutputhasbeenxedvia
digitalcommunication.
182MicroMotion9739MVDTransmitters