EasyManua.ls Logo

Emerson Rosemount 2460 System Hub - 6.24 Replacing the backup battery

Emerson Rosemount 2460 System Hub
240 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
145
Reference Manual
00809-0100-2460, Rev DE
Service and Troubleshooting
October 2018
Service and Troubleshooting
6.24 Replacing the backup battery
In case a battery warning appears, you will have to change the backup battery which is located on the
Main Board. To change the battery you will have to remove the Terminal Board which is stacked on top of
the Main Board.
Make sure that a button battery of type 3V CR 1632 lithium is used.
To replace the Backup Battery:
1. Turn off power supply by disconnecting external switch.
2. Remove the Terminal Board as described in “Exchanging the terminal board” on page 110.
3. Locate the Backup battery on the Main Board.
4. Replace the battery with a button battery of type 3V CR 1632 lithium. Ensure that the battery is
placed with the + side facing up.
5. Replace the Terminal Board.
6. Set date and time (see “Setting the real-time clock” on page 146).