Commissioning Proline t-mass I 300 HART
104 Endress+Hauser
Totalizer operation mode
→  104
Failure mode
→  104
Assign gas
→  104
Parameter overview with brief description
Parameter Prerequisite Description Selection Factory setting
Assign process variable Select process variable for
totalizer.
Off
Mass flow
Corrected volume
flow
FAD volume flow
*
Volume flow
Energy flow
*
Heat flow
*
Unit totalizer 1 to n A process variable is selected
in the Assign process variable
parameter (→  104) of the
Totalizer 1 to n submenu.
Select process variable totalizer
unit.
Unit choose list
Totalizer operation mode A process variable is selected
in the Assign process variable
parameter (→  104) of the
Totalizer 1 to n submenu.
Select totalizer calculation
mode.
Net flow total
Forward flow total
Reverse flow total
Failure mode A process variable is selected
in the Assign process variable
parameter (→  104) of the
Totalizer 1 to n submenu.
Define totalizer behavior in
alarm condition.
Stop
Actual value
Last valid value
Assign gas (Only with order code for
"Application package", option EV
"Second gas group")
Select the gas that the totalizer
uses. This gas is only totalized
when it is currently active
(‘Active gas’ parameter).
Both gases
Gas
Second gas
Both gases option
(only with order
code for
"Application
package", option
EV "Second gas
group")
Gas
* Visibility depends on order options or device settings