Commissioning Proline t-mass I 300 HART
104 Endress+Hauser
Totalizer operation mode
→ 104
Failure mode
→ 104
Assign gas
→ 104
Parameter overview with brief description
Parameter Prerequisite Description Selection Factory setting
Assign process variable – Select process variable for
totalizer.
• Off
• Mass flow
• Corrected volume
flow
• FAD volume flow
*
• Volume flow
• Energy flow
*
• Heat flow
*
–
Unit totalizer 1 to n A process variable is selected
in the Assign process variable
parameter (→ 104) of the
Totalizer 1 to n submenu.
Select process variable totalizer
unit.
Unit choose list –
Totalizer operation mode A process variable is selected
in the Assign process variable
parameter (→ 104) of the
Totalizer 1 to n submenu.
Select totalizer calculation
mode.
• Net flow total
• Forward flow total
• Reverse flow total
–
Failure mode A process variable is selected
in the Assign process variable
parameter (→ 104) of the
Totalizer 1 to n submenu.
Define totalizer behavior in
alarm condition.
• Stop
• Actual value
• Last valid value
–
Assign gas (Only with order code for
"Application package", option EV
"Second gas group")
– Select the gas that the totalizer
uses. This gas is only totalized
when it is currently active
(‘Active gas’ parameter).
• Both gases
• Gas
• Second gas
• Both gases option
(only with order
code for
"Application
package", option
EV "Second gas
group")
• Gas
* Visibility depends on order options or device settings