Adjustment values in use
→ 117
Parameter overview with brief description
Parameter Description Selection / User interface /
User entry
Factory setting
Activate in-situ adjustment Activate the in-situ adjustment. The points
stored by the user are used for the in-situ
adjustment.
• No
• Yes
–
Input type reference value Select input type for the reference value. • Off
• Manual
• Current input 1
*
• Current input 2
*
• Current input 3
*
• External value
*
–
Delete values Delete previous adjustment values and
descriptions.
• No
• Yes
–
Confirm Confirm deletion. • No
• Yes
–
Select flow reference Select process variable. This process variable
is used as reference value for the in situ
adjustment.
• Mass flow
• Corrected volume flow
• FAD volume flow
*
• Volume flow
–
Stability check Activate stability check. New adjustment
value is only accepted when the
measurement is stable.
• No
• Yes
–
Actual flow value Shows the actual flow in relation to the
maximum, factory-measured value that is
adapted to the actual process conditions.
–2 000 to 2 000 % –
External reference value Shows the external reference value for the in
situ adjustment.
Signed floating-point number –
Reference value Enter fixed value as reference value used for
the in situ adjustment.
Signed floating-point number –
Apply value Apply the actual value. • No
• Yes
–
Status Shows the validity of the actual reference
value.
• Passed
• Replaced
• Unstable
• Invalid
–
Description 1 Description for in-situ adjustment: e.g.
facility, operator, date.
- -
Description 2 Description for in-situ adjustment: e.g.
facility, operator, date.
- -
Description 3 Description for in-situ adjustment: e.g.
facility, operator, date.
- -
Description 4 Description for in-situ adjustment: e.g.
facility, operator, date.
- -