Proline t-mass I 300 HART Commissioning
Endress+Hauser 121
Pulse output simulation 1 to n
→ 122
Pulse value 1 to n
→ 122
Switch output simulation 1 to n
→ 122
Switch status 1 to n
→ 122
Relay output 1 to n simulation
→ 122
Switch status 1 to n
→ 122
Device alarm simulation
→ 122
Diagnostic event category
→ 122
Diagnostic event simulation
→ 122
Parameter overview with brief description
Parameter Prerequisite Description Selection / User entry
Assign simulation process variable – Select a process variable for the
simulation process that is activated.
• Off
• Temperature
• Mass flow
• Corrected volume flow
• FAD volume flow
*
• Volume flow
• Energy flow
*
• Heat flow
*
• Density
• Flow velocity
Process variable value A process variable is selected in the
Assign simulation process variable
parameter (→ 121).
Enter the simulation value for the
selected process variable.
Depends on the process
variable selected
Current input 1 to n simulation – Switch simulation of the current input
on and off.
• Off
• On
Value current input 1 to n In the Current input 1 to n simulation
parameter, the On option is selected.
Enter the current value for simulation. 0 to 22.5 mA
Status input simulation 1 to n – Switch simulation of the status input on
and off.
• Off
• On
Input signal level 1 to n In the Status input simulation
parameter, the On option is selected.
Select the signal level for the simulation
of the status input.
• High
• Low
Current output 1 to n simulation – Switch the simulation of the current
output on and off.
• Off
• On
Value current output 1 to n In the Current output 1 to n
simulation parameter, the On option is
selected.
Enter the current value for simulation. 3.59 to 22.5 mA
Frequency output simulation 1 to n In the Operating mode parameter, the
Frequency option is selected.
Switch the simulation of the frequency
output on and off.
• Off
• On
Frequency value 1 to n In the Frequency output simulation
1 to n parameter, the On option is
selected.
Enter the frequency value for the
simulation.
0.0 to 12 500.0 Hz