EasyManua.ls Logo

Endress+Hauser HART Proline t-mass I 300 - Page 134

Endress+Hauser HART Proline t-mass I 300
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Proline t-mass I 300 HART
134 Endress+Hauser
Parameter overview with brief description
Parameter Prerequisite Description Selection / User entry / 
User interface
Assign channel 1 The Extended HistoROM application
package is available.
Assign process variable to logging
channel.
Off
Temperature
Mass flow
Corrected volume flow
FAD volume flow
*
Volume flow
Energy flow
Heat flow
*
Density
Flow velocity
Pressure
2nd temperature delta
heat
*
Electronic temperature
Current output 1
*
Current output 2
*
Current output 3
*
Current output 4
*
Assign channel 2 The Extended HistoROM application
package is available.
The software options currently
enabled are displayed in the
Software option overview
parameter.
Assign process variable to logging
channel.
For the picklist, see the
Assign channel 1
parameter (→  134)
Assign channel 3 The Extended HistoROM application
package is available.
The software options currently
enabled are displayed in the
Software option overview
parameter.
Assign process variable to logging
channel.
For the picklist, see the
Assign channel 1
parameter (→  134)
Assign channel 4 The Extended HistoROM application
package is available.
The software options currently
enabled are displayed in the
Software option overview
parameter.
Assign process variable to logging
channel.
For the picklist, see the
Assign channel 1
parameter (→  134)
Logging interval The Extended HistoROM application
package is available.
Define the logging interval for data
logging. This value defines the time
interval between the individual data
points in the memory.
0.1 to 3 600.0 s
Clear logging data The Extended HistoROM application
package is available.
Clear the entire logging data. Cancel
Clear data
Data logging Select the data logging method. Overwriting
Not overwriting
Logging delay In the Data logging parameter, the Not
overwriting option is selected.
Enter the time delay for measured
value logging.
0 to 999 h
Data logging control In the Data logging parameter, the Not
overwriting option is selected.
Start and stop measured value logging. None
Delete + start
Stop
Data logging status In the Data logging parameter, the Not
overwriting option is selected.
Displays the measured value logging
status.
Done
Delay active
Active
Stopped
Entire logging duration In the Data logging parameter, the Not
overwriting option is selected.
Displays the total logging duration. Positive floating-point
number
* Visibility depends on order options or device settings

Table of Contents

Related product manuals