Proline t-mass I 300 HART Technical data
Endress+Hauser 173
Signal on alarm Depending on the interface, failure information is displayed as follows:
Current output 0/4 to 20 mA
4 to 20 mA
Failure mode Choose from:
4 to 20 mA in accordance with NAMUR recommendation NE 43
4 to 20 mA in accordance with US
Min. value: 3.59 mA
Max. value: 22.5 mA
Freely definable value between: 3.59 to 22.5 mA
Actual value
Last valid value
0 to 20 mA
Failure mode Choose from:
Maximum alarm: 22 mA
Freely definable value between: 0 to 20.5 mA
Pulse/frequency/switch output
Pulse output
Failure mode Choose from:
Actual value
No pulses
Frequency output
Failure mode Choose from:
Actual value
0 Hz
Defined value (f
max
2 to 12 500 Hz)
Switch output
Failure mode Choose from:
Current status
Open
Closed
Relay output
Failure mode Choose from:
Current status
Open
Closed
Local display
Plain text display With information on cause and remedial measures
Backlight Red backlighting indicates a device error.
Status signal as per NAMUR recommendation NE 107