EasyManua.ls Logo

Endress+Hauser HART Proline t-mass I 300 - Page 172

Endress+Hauser HART Proline t-mass I 300
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical data Proline t-mass I 300 HART
172 Endress+Hauser
Number of switching
cycles
Unlimited
Assignable functions Off
On
Diagnostic behavior
Limit value
Off
Mass flow
Volume flow
Corrected volume flow
FAD volume flow
Heat flow
Energy flow
Flow velocity
Density
Calorific value
Temperature
2nd temperature heat difference
Totalizer 1-3
Electronics temperature
Flow direction monitoring
Status
Low flow cut off
Relay output
Function Switch output
Version Relay output, galvanically isolated
Switching behavior Can be set to:
NO (normally open), factory setting
NC (normally closed)
Maximum switching
capacity (passive)
DC 30 V, 0.1 A
AC 30 V, 0.5 A
Assignable functions Off
On
Diagnostic behavior
Limit value
Off
Mass flow
Volume flow
Corrected volume flow
FAD volume flow
Heat flow
Energy flow
Flow velocity
Density
Temperature
2nd temperature heat difference
Totalizer 1-3
Electronics temperature
Flow direction monitoring
Status
Low flow cut off
User-configurable input/output
One specific input or output is assigned to a user-configurable input/output (configurable
I/O) during device commissioning.
The following inputs and outputs are available for assignment:
Choice of current output: 4 to 20 mA (active), 0/4 to 20 mA (passive)
Pulse/frequency/switch output
Choice of current input: 4 to 20 mA (active), 0/4 to 20 mA (passive)
Status input

Table of Contents

Related product manuals