Proline t-mass I 300 HART Technical data
Endress+Hauser 171
Open-circuit voltage DC 28.8 V (active)
Voltage drop For 22.5 mA: ≤ DC 2 V
Pulse output
Maximum input values DC 30 V, 250 mA (passive)
Maximum output current 22.5 mA (active)
Open-circuit voltage DC 28.8 V (active)
Pulse width Configurable: 0.05 to 2 000 ms
Maximum pulse rate 10 000 Impulse/s
Pulse value Adjustable
Assignable measured
variables
Mass flow
Volume flow
Corrected volume flow
FAD volume flow
Energy flow
Heat flow
For SIL (application package), only mass flow
Frequency output
Maximum input values DC 30 V, 250 mA (passive)
Maximum output current 22.5 mA (active)
Open-circuit voltage DC 28.8 V (active)
Output frequency Adjustable: end value frequency 2 to 10 000 Hz (f
max
= 12 500 Hz)
Damping Configurable: 0 to 999.9 s
Pulse/pause ratio 1:1
Assignable measured
variables
Mass flow
Volume flow
Corrected volume flow
FAD volume flow
Flow velocity
Temperature
Energy flow
Pressure
Density
Heat flow
Electronics temperature
2nd temperature heat difference
For SIL (application package), only mass flow
Switch output
Maximum input values DC 30 V, 250 mA (passive)
Open-circuit voltage DC 28.8 V (active)
Switching behavior Binary, conductive or non-conductive
Switching delay Configurable: 0 to 100 s