Installation Proline t-mass I 300 HART
22 Endress+Hauser
20 × DN 2 × DN
A0040193
 4 Reduction
30 × DN 2 × DN
A0040192
 5 Expansion
20 × DN 2 × DN
A0039440
 6 90° elbow
25 × DN 2 × DN
A0039441
 7 2 × 90° elbow
50 × DN 2 × DN
A0039445
 8 Control valve
35 × DN 2 × DN
A0039442
 9 2 × 90 ° elbow, 3-dimensional
%
DN
40353025
20
15
10
5
0
1
2
3
4
5
6
7
A
B
1
2
3
4
A0045846
 10 The additional measured error to be expected without flow conditioners depending on the disturbance
type and inlet run
A Additional measured error (%)
B Inlet run (DN)
1 2 × 90 ° elbow, 3-dimensional
2 Expansion
3 2 × 90° elbow
4 Reduction or 90° elbow
Flow conditioner
Use flow conditioners if it is not possible to observe the required inlet runs. Flow
conditioners improve the flow profile and therefore reduce the necessary inlet runs.