Proline t-mass I 300 HART Installation
Endress+Hauser 23
Mount the flow conditioner in flow direction in front of the measuring device.
Available in the following flange standards:
ASME B16.5 Cl. 150/Cl. 300
EN 1092-1 PN10/PN16/PN25/PN40
JIS B2220 10K/20K
Available in the following line sizes:
DN 80 (3")
DN 100 (4")
DN 150 (6")
DN 200 (8")
DN 250 (10")
DN 300 (12")
1 22
A0039538
1 Flow conditioner
2 Seal
10 × DN
5 × DN
1
5 × DN
A0039424
 11 Recommended inlet and outlet runs when using a flow conditioner
1 Flow conditioner
In the case of bidirectional sensors, also observe the inlet run in the opposite
direction.