EasyManua.ls Logo

Endress+Hauser Levelflex FMP55 - Level Correction; Manual Thickness Upper Layer

Endress+Hauser Levelflex FMP55
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating menu Levelflex FMP55 HART
144 Endress+Hauser
1
2
UB
UB
A0013220
1 Suppression of interference echoes at the top end of the probe.
2 Suppression of the level signal in case of a flooded bypass.
UB Upper blocking distance
Level correction
Navigation  Setup → Advanced setup → Interface → Level correction
Description Specify level correction (if required).
User entry –200 000.0 to 200 000.0 %
Factory setting 0.0 %
Additional information The value specified in this parameter is added to the measured total and interface levels
(before linearization).
Manual thickness upper layer
Navigation Setup → Advanced setup → Interface → Manual thickness upper layer
Description Specify the manually determined interface thickness UP (i.e. the thickness of the upper
medium).
User entry 0 to 200 m
Factory setting 0 m

Table of Contents

Other manuals for Endress+Hauser Levelflex FMP55