EasyManua.ls Logo

Endress+Hauser Levelflex FMP55 - Connecting the Measuring Device

Endress+Hauser Levelflex FMP55
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electrical connection Levelflex FMP55 HART
48 Endress+Hauser
4-wire, 4-20mA HART, active
"Power supply; Output"
1)
Terminal voltage Maximum load R
max
K: 4-wire 90-253VAC; 4-20mA HART 90 to 253 V
AC
(50 to 60 Hz), overvoltage
category II
500 Ω
L: 4-wire 10,4-48VDC; 4-20mA HART 10.4 to 48 V
DC
1) Feature 020 of the product structure
7.1.5 Overvoltage protection
If the measuring device is used for level measurement in flammable liquids which requires
the use of overvoltage protection according to DIN EN 60079-14, standard for
test procedures 60060-1 (10 kA, pulse 8/20 μs), an overvoltage protection module has to
be installed.
Integrated overvoltage protection module
An integrated overvoltage protection module is available for 2-wire HART as well as
PROFIBUS PA and FOUNDATION Fieldbus devices.
Product structure: Feature 610 "Accessory mounted", option NA "Overvoltage protection".
Technical data
Resistance per channel 2 × 0.5 Ω max.
Threshold DC voltage 400 to 700 V
Threshold impulse voltage < 800 V
Capacitance at 1 MHz < 1.5 pF
Nominal arrest impulse voltage (8/20 μs) 10 kA
External overvoltage protection module
HAW562 or HAW569 from Endress+Hauser are suited as external overvoltage protection.
For detailed information please refer to the following documents:
HAW562: TI01012K
HAW569: TI01013K
7.2 Connecting the measuring device
L
WARNING
Risk of explosion!
Observe applicable national standards.
Comply with the specifications in the Safety Instructions (XA).
Use specified cable glands only.
Check to ensure that the power supply matches the information on the nameplate.
Switch off the power supply before connecting the device.
Connect the potential matching line to the outer ground terminal before applying the
power supply.
Required tools/accessories:
For devices with a cover lock: Allen key AF3
Wire stripper
When using stranded cables: One ferrule for every wire to be connected.

Table of Contents