Operating menu Levelflex FMP55 HART
182 Endress+Hauser
Selection • x
• x.x
• x.xx
• x.xxx
• x.xxxx
Factory setting x.xxxx
Additional information • Is only valid for numbers in the operating menu (e.g. Empty calibration, Full
calibration), but not for the measured value display. The number of decimal places for
the measured value display is defined in the Decimal places 1 to 4 → 179
parameters.
• The setting does not affect the accuracy of the measurement or the calculations.
Backlight
Navigation Setup → Advanced setup → Display → Backlight
Prerequisite The device has the SD03 local display (with optical keys).
Description Switch the local display backlight on and off.
Selection • Disable
• Enable
Factory setting Disable
Additional information Meaning of the options
• Disable
Switches the backlight off.
• Enable
Switches the backlight on.
Regardless of the setting in this parameter the backlight may be automatically
switched off by the device if the supply voltage is too low.
Contrast display
Navigation Setup → Advanced setup → Display → Contrast display
Description Adjust local display contrast setting to ambient conditions (e.g. lighting or reading angle).
User entry 20 to 80 %
Factory setting Dependent on the display.
Additional information Setting the contrast via push-buttons:
• Darker: press the
buttons simultaneously.
• Brighter: press the
buttons simultaneously.