Levelflex FMP55 HART Operating menu
Endress+Hauser 183
"Configuration backup display" submenu
This submenu is only visible if a display module is connected to the device.
The configuration of the device can be saved to the display module at a certain point of
time (backup). The saved configurateion can be restored to the device if required, e.g. in
order to bring the device back into a defined state. The configuration can also be
transferred to a different device of the same type using the display module.
Configurations can only be exchanged between devices which are in the same
operating mode (see the Operating mode parameter (→ 128)).
Navigation Setup → Advanced setup → Configuration backup display
Operating time
Navigation Setup → Advanced setup → Configuration backup display → Operating time
Description Indicates how long the device has been in operation.
Additional information Maximum time
9 999 d ( ≈ 27 years)
Last backup
Navigation Setup → Advanced setup → Configuration backup display → Last backup
Description Indicates when the last data backup was saved to the display module.
Configuration management
Navigation Setup → Advanced setup → Configuration backup display → Configuration
management
Description Select action for managing the device data in the display module.
Selection • Cancel
• Execute backup
• Restore
• Duplicate
• Compare
• Clear backup data
Factory setting Cancel