ProlineйPromagйHй100йEtherNet/IP Operationйoptions
Endress+Hauser 39
8.3.2 Prerequisites
Hardware
Connectingйcable StandardйEthernetйcableйwithйRJ45йconnector
Computer RJ45йinterface
Measuringйdevice: Webйserverйmustйbeйenabled;йfactoryйsetting:йON
ForйinformationйonйenablingйtheйWebйserverйсййй42)
IPйaddress IfйtheйIPйaddressйofйtheйdeviceйisйnotйknown,йcommunicationйwithйtheйWeb
serverйcanйbeйestablishedйviaйtheйstandardйIPйaddressй192.168.1.212.
TheйDHCPйfunctionйisйenabledйinйtheйdeviceйatйtheйfactory,йi.e.йtheйdevice
expectsйanйIPйaddressйtoйbeйassignedйbyйtheйnetwork.йThisйfunctionйcanйbe
disabledйandйtheйdeviceйcanйbeйsetйtoйtheйstandardйIPйaddressй192.168.1.212:
setйswitchйDIPйswitchйNo.й10йfromйOFFййON.
OFF ON
1
2
3
4
5
6
7
8
9
10
1
2
4
8
16
32
64
128
- Write protection
- Default Ethernet
network settings
IP 192.168.1.212
IP Address setting
(last octet)
ййA0017965
• OnceйtheйDIPйswitchйhasйbeenйactivated,йtheйdeviceйmustйbeйrestarted
beforeйtheйdeviceйusesйtheйstandardйIPйaddress.
• IfйtheйstandardйIPйaddressйсDIPйswitchйNo.й10й=йON)йisйused,йthereйis
noйconnectionйtoйtheйEtherNet/IPйnetwork.
Software of the computer
Webйbrowsersйsupported • MicrosoftйInternetйExplorerйсmin.й8.x)
• MozillaйFirefox
• Googleйchrome
Recommendedйoperating
systems
• WindowsйXP
• Windowsй7
UserйrightsйforйTCP/IPйsettings UserйrightsйrequiredйforйTCP/IPйsettingsйсe.g.йforйchangesйtoйIPйaddress,йsubnet
mask)
Computerйconfiguration • JavaScriptйisйenabled
• IfйJavaScriptйcannotйbeйenabled,йenterйhttp://XXX.XXX.X.XXX/basic.htmlйin
theйaddressйlineйofйtheйWebйbrowser,йe.g.йhttp://192.168.1.212/basic.html.
Aйfullyйfunctionalйbutйsimplifiedйversionйofйtheйoperatingйmenuйstructure
startsйinйtheйWebйbrowser.
Whenйinstallingйaйnewйfirmwareйversion:
Toйenableйcorrectйdataйdisplay,йclearйtheйtemporaryйmemoryйсcache)йofйtheйWeb
browserйunderйInternet options.
8.3.3 Establishing a connection
Configuring the Internet protocol of the computer
TheйfollowingйinformationйrefersйtoйtheйdefaultйEthernetйsettingsйofйtheйdevice.