Technicalйdata ProlineйPromagйHй100йEtherNet/IP
90 Endress+Hauser
Configuration
ConfigurationйAssembly Onlyйtheйmostйcommonйconfigurationsйareйlistedйbelow.
• Softwareйwriteйprotection
• Massйflowйunit
• Massйunit
• Volumeйflowйunit
• Volumeйunit
• Correctedйvolumeйflowйunit
• Correctedйvolumeйunit
• Densityйunit
• Referenceйdensityйunit
• Temperatureйunit
• Pressureйunit
• Length
• Totalizerй1-3:
– Assignment
– Unit
– Measuringйmode
– Failureйmode
• Alarmйdelay
16.5 Power supply
Terminalйassignment сййй28)
Pinйassignment,йdeviceйplug сййй29)
Supplyйvoltage Transmitter
Forйdeviceйversionйwithйallйcommunicationйtypes:йDCй20йtoй30йV
Theйpowerйunitйmustйbeйtestedйtoйensureйitйmeetsйsafetyйrequirementsйсe.g.йPELV,йSELV).
Powerйconsumption Transmitter
Order code for
"Output"
Maximum
Power consumption
OptionйN:йEtherNet/IP 3.5йW
Currentйconsumption Transmitter
Order code for
"Output"
Maximum
Current consumption
Maximum
switch-on current
OptionйN:йEtherNet/IP 145йmA 18йAйс<0.125йms)
Powerйsupplyйfailure • Totalizersйstopйatйtheйlastйvalueйmeasured.
• Dependingйonйtheйdeviceйversion,йtheйconfigurationйisйretainedйinйtheйdeviceйmemoryйor
inйtheйplug-inйmemoryйсHistoROMйDAT).
• Errorйmessagesйсincl.йtotalйoperatedйhours)йareйstored.
Electricalйconnection сййй29)
Potentialйequalization сййй31)