ProlineйPromagйHй100йEtherNet/IP Technicalйdata
Endress+Hauser 85
16 Technical data
16.1 Application
Dependingйonйtheйversionйordered,йtheйmeasuringйdeviceйcanйalsoйmeasureйpotentially
explosive,йflammable,йpoisonousйandйoxidizingйmedia.
Toйensureйthatйtheйdeviceйremainsйinйproperйoperatingйconditionйforйitsйserviceйlife,йuseйthe
measuringйdeviceйonlyйforйmediaйagainstйwhichйtheйprocess-wettedйmaterialsйare
adequatelyйresistant.
16.2 Function and system design
Measuringйprinciple ElectromagneticйflowйmeasurementйonйtheйbasisйofйFaraday's law of magnetic induction.
Measuringйsystem Oneйdeviceйversionйisйavailable:йcompactйversion,йtransmitterйandйsensorйformйaйmechanical
unit.
Forйinformationйonйtheйstructureйofйtheйdeviceйсййй11)
16.3 Input
Measuredйvariable Direct measured variables
Volumeйflowйсproportionalйtoйinducedйvoltage)
TemperatureйсDNй15йtoй150йс½йtoй6"))
Electricalйconductivity
Calculated measured variables
Massйflow
Correctedйvolumeйflow
Correctedйelectricalйconductivity
Measuringйrange Typicallyйvй=й0.01йtoй10йm/sйс0.03йtoй33йft/s)йwithйtheйspecifiedйaccuracy
Electricalйconductivity:й5йtoй10й000йµS/cm/cm
Flow characteristic values in SI units
Nominal
diameter
Recommended flow Factory settings
min./max. full scale value
(v ~ 0.3/10 m/s)
Low flow cut off
(v ~ 0.04 m/s)
[mm] [in] [dm
3
/min] [dm
3
/min]
2 1/12 0.06йtoй1.8 0.01
4 1/8 0.25йtoй7 0.05
8 3/8 1йtoй30 0.1
15 ½ 4йtoй100 0.5
25 1 9йtoй300 1
40 1й½ 25йtoй700 3
50 2 35йtoй1й100 5