Working area
Dependingйonйtheйselectedйfunctionйandйtheйrelatedйsubmenus,йvariousйactionsйcanйbe
performedйinйthisйarea:
• Configuringйparameters
• Readingйmeasuredйvalues
• Callingйupйhelpйtext
• Startingйanйupload/download
8.3.6 Disabling the Web server
TheйWebйserverйforйtheйmeasuringйdeviceйcanйenabledйandйdisabledйasйrequiredйviaйthe
Web server functionalityйparameter.
Navigation
"Expert"йmenuййCommunicationййWebйserver
Parameter overview with brief description
Parameter Description Selection Factory setting
Webйserverйfunctionality SwitchйtheйWebйserverйonйandйoff. • Off
• On
On
Enabling the Web server
IfйtheйWebйserverйisйdisabledйitйcanйonlyйbeйre-enabledйwithйtheйWeb server functionality
parameterйviaйtheйfollowingйoperatingйoptions:
Viaй"FieldCare"йoperatingйtool
8.3.7 Logging out
Beforeйloggingйout,йperformйaйdataйbackupйviaйtheйData managementйfunction
сuploadйconfigurationйfromйdevice)йifйnecessary.
1. SelectйtheйLogoutйentryйinйtheйfunctionйrow.
TheйhomeйpageйwithйtheйLoginйboxйappears.
2. CloseйtheйWebйbrowser.
3. ResetйtheйmodifiedйpropertiesйofйtheйInternetйprotocolйсTCP/IP)йifйtheyйareйnoйlonger
neededйсййй39).
IfйcommunicationйwithйtheйWebйserverйwasйestablishedйviaйtheйstandardйIPйaddress
192.168.1.212,йDIPйswitchйNo.й10йmustйbeйresetйсfromйONййOFF)йandйtheйIPйaddress
ofйtheйdeviceйisйactiveйagainйforйnetworkйcommunication.