ProlineйPromagйHй100йEtherNet/IP Operation
Endress+Hauser 65
11.3.2 Totalizer
Theй"Totalizer" submenuйcontainsйallйtheйparametersйneededйtoйdisplayйtheйcurrent
measuredйvaluesйforйeveryйtotalizer.
Navigation
"Diagnostics"йmenuййMeasuredйvaluesййTotalizer
Structure of the submenu
Totalizer
Totalizerйvalue
Totalizerйoverflow
Parameter overview with brief description
Parameter Description User interface Factory setting
Totalizerйvalueй# Displaysйtheйcurrentйtotalizerйcounterйvalue. Signedйfloating-pointйnumber 0йl
Totalizerйoverflowй# Displaysйtheйcurrentйtotalizerйoverflow. –32й000.0йtoй32й000.0 0
11.4 Adapting the measuring device to the process
conditions
Theйfollowingйareйavailableйforйthisйpurpose:
• BasicйsettingsйusingйtheйSetupйmenuсййй48)
• AdvancedйsettingsйusingйtheйAdvanced setupйsubmenuсййй55)
11.5 Performing a totalizer reset
Function scope of "Control Totalizer " parameter
Options Description
Totalize Theйtotalizerйisйstarted.
Stop Totalizingйisйstopped.
Resetй+йhold Theйtotalingйprocessйisйstoppedйandйtheйtotalizerйisйresetйtoй0.
Presetй+йhold Theйtotalingйprocessйisйstoppedйandйtheйtotalizerйisйsetйtoйitsйdefinedйstartйvalue
fromйtheйPreset value parameter.
Resetй+йtotalize Theйtotalizerйisйresetйtoй0йandйtheйtotalingйprocessйisйrestarted.
Presetй+йtotalize TheйtotalizerйisйsetйtoйtheйdefinedйstartйvalueйinйPreset value parameterandйthe
totalingйprocessйisйrestarted.
Function scope of "Reset all totalizers" parameter
Options Description
Resetй+йtotalize Resetsйallйtotalizersйtoй0йandйrestartsйtheйtotalingйprocess.йThisйdeletesйallйtheйflow
valuesйpreviouslyйtotalized.
Navigation
"Operation"йmenuййOperation