10 Block SPECIAL FUNCTION Device Functions Proline Promass 83 MODBUS RS485
128 Endress+Hauser
CONCENTRATION
SELECTOR
MODBUS register:
Data type:
Access:
7022
2291
Integer
read/write
For selecting the concentration specification. There are four different
specifications available, via which various concentrations can be defined.
Options:
0 = CONC. # 1
1 = CONC. # 2
2 = CONC. # 3
3 = CONC. # 4
Factory setting: CONC. # 1
! Note!
A specific name can be given to the batching specification (CONC. # 1 to 4) in
the CONCENTRATION NAME function (7031).
By selecting a concentration specification and the (subsequent) relevant
settings, up to 4 different concentrations can be preconfigured and selected
when needed.
All settings in the subsequent functions of this function group are each only
valid for the concentration specification selected in the function
CONCENTRATION SELECTOR (7022). In other words, the entry or option is
assigned to the concentration specification currently selected (e.g. in the factory
setting CONC. # 1).
CONCENTRATION
NAME
MODBUS register:
Data type:
Access:
7031
See Note
String (8)
read/write
For entering a specific name for the concentration specification selected.
User input: max. 8-character text, permissible: A-Z, 0-9
Factory setting: Name of concentration specification (depends on selection in
the function CONCENTRATION SELECTOR (7022), e.g. “CONC. # 1”).
! Note for MODBUS!
The MOBBUS register depends on the option selected in the function
CONCENTRATION SELECTOR (7022):
CONC. # 1 = MODBUS register 2292
CONC. # 2 = MODBUS register 2296
CONC. # 3 = MODBUS register 2300
CONC. # 4 = MODBUS register 2304
COEFFICIENT A0
MODBUS register:
Data type:
Access:
7032
See Note
Float
read/write
Coefficient A0 entry.
User input: 5-digit floating-point number
Factory setting: 0
! Note for MODBUS!
The MOBBUS register depends on the option selected in the function
CONCENTRATION SELECTOR (7022):
CONC. # 1 = MODBUS register 2227
CONC. # 2 = MODBUS register 2229
CONC. # 3 = MODBUS register 2231
CONC. # 4 = MODBUS register 2233
COEFFICIENT A1
MODBUS register:
Data type:
Access:
7033
See Note
Float
read/write
Coefficient A1 entry.
User input: 5-digit floating-point number
Factory setting: 0
! Note for MODBUS!
The MOBBUS register depends on the option selected in the function
CONCENTRATION SELECTOR (7022):
CONC. # 1 = MODBUS register 2235
CONC. # 2 = MODBUS register 2237
CONC. # 3 = MODBUS register 2239
CONC. # 4 = MODBUS register 2241
Function description
SPECIAL FUNCTION DENSITY FUNCTIONS CONFIGURATION