Device Functions Proline Promass 83 MODBUS RS485 10 Block SPECIAL FUNCTION
Endress+Hauser 135
CALCULATION
MODE
MODBUS register:
Data type:
Access:
7206
See Note
Integer
read/write
! Note!
Function is only available if MODE 1 or MODE 2 is selected in the function
COMPENSATION MODE (7205).
For selecting the method for calculating the recorded after run quantities.
Options:
0 = ALL
All after run quantities are used in the calculation.
1 = SELECTION
The after run quantities recorded are filtered. The smallest and largest after run
quantity are not taken into account in the calculation (extreme value filter).
Factory setting: ALL
! Note!
Machine-related (larger) “extreme values”, occurring especially at startup, delay
correction and distort real reproducibility. By selecting “SELECT”, these “extreme
values” are not taken into account.
! Note for MODBUS!
The MOBBUS register depends on the option selected in the function BATCH
SELECTOR (7200):
• BATCH # 1 = MODBUS register 6362
• BATCH # 2 = MODBUS register 6363
• BATCH # 3 = MODBUS register 6364
• BATCH # 4 = MODBUS register 6365
• BATCH # 5 = MODBUS register 6366
• BATCH # 6 = MODBUS register 6367
Example:
Function CALCULATION MODE (7206) = SELECTION
Function AVERAGING DRIP (7207) = 5
From five after run quantities recorded, the largest and smallest are not used.
From the remaining three after run quantities, an average after run quantity is
calculated which is taken into account at the next batching.
Function description
SPECIAL FUNCTION → BATCHING FUNCTION → CONFIGURATION