3 Transducer Blocks (transmission blocks) PROline Prosonic Flow 93 FOUNDATION Fieldbus
158 Endress+Hauser
System -
Trouble-
shooting
AUTO - OOS Use this parameter to rectify errors in the EEPROM. The EEPROM
is divided into a number of blocks. For fault elimination, select the
block in question and acknowledge the error.
Caution!
When you clear an error in a block, the parameters of the block
you select are reset to their factory settings.
Options:
CANCEL
MEASURED VALUES
SYSTEM UNITS
DENSITY PARAMETERS
QUICK SETUP
USER INTERFACE
TOTALIZER
COMMUNICATION
PROCESS PARAMETER
SYSTEM PARAMETER
SENSOR DATA
BATCH FUNCTION
ADVANCED DIAGNOSIS
AMPLIFIER PARAMETERS
SUPERVISION
VERSION INFO
SERVICE & ANALYSIS
PRODUCTION INFO
FILTER PARAMETER
Factory setting:
CANCEL
System -
Operation
Time
read Displays the entire operation time (in seconds) since the flowmeter
was commissioned.
T-DAT Save/
Load
AUTO - OOS Use this parameter to save the parameter settings / configuration
of the transmitter in a transmitter DAT (T-DAT), or to load the
parameter settings from the T-DAT into the EEPROM.
Application examples:
After commissioning, the current measuring point parameters
can be saved to the T-DAT as a backup.
If the transmitter is replaced for some reason, the data from the
T-DAT can be loaded into the new transmitter (EEPROM).
Options:
CANCEL
SAVE (from EEPROM to T-DAT)
LOAD (from T-DAT to EEPROM)
Factory setting:
CANCEL
Note!
If the power supply fails, the totalizer readings are automatically
saved to the EEPROM.
Amp. Device
Family
read only This parameter is only used in the event of servicing.
TRANSDUCER_DIAG (E+H parameters) / base index 1600
Parameter Write access
with operating
mode
(MODE_BLK)
Description

Table of Contents

Other manuals for Endress+Hauser Proline Prosonic Flow 93

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Proline Prosonic Flow 93 and is the answer not in the manual?

Endress+Hauser Proline Prosonic Flow 93 Specifications

General IconGeneral
BrandEndress+Hauser
ModelProline Prosonic Flow 93
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals