5 Block USER INTERFACE PROline Prosonic Flow 93 FOUNDATION Fieldbus
36 Endress+Hauser
Contd:
ASSIGN
(2420)
Note!
Multiplex mode is suspended as soon as a fault/notice message is generated.
The appropriate error message appears on the display:
Fault message (identified by a lightning icon): Multiplex mode is resumed as
soon as the fault is no longer active.
Notice message (identified by an exclamation mark): Multiplex mode is
resumed as soon as the notice message is no longer active.
If a channel is not visible, it does not appear in the options. The channels can
be displayed or hidden by means of the MEASUREMENT (6880) function.
100% VALUE
(2421)
Note!
This function is not available unless one of the following options was selected
in the ASSIGN (2420) function:
VOLUME FLOW % CH1
VOLUME FLOW % CH2
CALCULATED VOLUME FLOW %
VOLUME FLOW BARGRAPH % CH1
VOLUME FLOW BARGRAPH % CH2
CALCULATED VOLUME FLOW BARGRAPH %
Use this function to define the flow value to be shown on the display as the
100% value.
User input:
5-digit floating-point number
Factory setting:
Depends on country, [value] / [dm³...m³ or US-gal...US-Mgal]
FORMAT
(2422)
Note!
This function is not available unless a numerical option was selected in the
ASSIGN (2420) function.
Use this function to define the maximum number of places after the decimal
point for the second value displayed in the additional line.
Options:
XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX
Factory setting:
X.XXXX
Note!
Note that this setting only affects the reading as it appears on the display, it
has no influence on the accuracy of the system's calculations.
The places after the decimal point as computed by the measuring device
cannot always be displayed, depending on this setting and the engineering
unit. In these instances an arrow appears on the display between the
measured value and the engineering unit, (e.g. 1.2
m³/h), indicating that
the measuring system is computing with more decimal places than can be
shown on the display.
Function description