6 Block TOTALIZERS PROline Prosonic Flow 93 FOUNDATION Fieldbus
46 Endress+Hauser
UNIT TOTALIZER
(3001)
Use this function to define the unit for the totalizer's measured variable, as
selected beforehand.
Options:
Metric:
Cubic centimetre
→ cm³
Cubic decimetre
→ dm³
Cubic metre
→ m³
Millilitre
→ ml
Litre
→ l
Hectolitre
→ hl
Megalitre
→ Ml MEGA
US:
Cubic centimetre
→ cc
Acre foot
→ af
Cubic foot
→ ft³
Fluid ounce
→ oz f
Gallon
→ US gal
Million gallon
→ US Mgal
Barrel (normal fluids: 31.5 gal/bbl)
→ US bbl NORM.FL.
Barrel (beer: 31.0 gal/bbl)
→ US bbl BEER
Barrel (petrochemicals: 42.0 gal/bbl)
→ US bbl PETROCH.
Barrel (filling tanks: 55.0 gal/bbl)
→ US bbl TANK
Imperial:
Gallon
→ imp. gal
Mega gallon
→ imp. Mgal
Barrel (beer: 36.0 gal/bbl)
→ imp. bbl BEER
Barrel (petrochemicals: 34.97 gal/bbl)
→ imp. bbl PETROCH.
Factory setting:
Depends on country (dm³...m³ or US gal...US Mgal),
corresponds to the totalizer unit factory setting (see Page 76 ff.).
Note!
The unit selected here has no effect on the FOUNDATION Fieldbus. It is only
used for the local display and for assigned instrument functions.
TOTALIZER MODE
(3002)
Use this function to define how the flow components are to be totalled by the
totalizer.
Options:
BALANCE
→ Positive and negative flow components.
(The positive and negative flow components are balanced. In other words, net
flow in the flow direction is registered.)
FORWARD
→ Positive flow components only.
REVERSE
→ Negative flow components only.
Factory setting:
Totalizer 1 = BALANCE
Totalizer 2 = F ORWA RD
Totalizer 3 = REV ERSE
RESET
TOTALIZER
(3003)
Use this function to reset the sum and the overflow of the totalizer to “0”.
Options:
NO
YES
Factory setting:
NO
Function description
TOTALIZERS → TOTALIZER (1...3) → CONFIGURATION