PROline Prosonic Flow 93 FOUNDATION Fieldbus 8 Block SUPERVISION
Endress+Hauser 71
8.1.2 Function group: OPERATION
SUPERVISION J
Þ
SYSTEM JAA
Þ
CONFIGURATION 800
ß
OPERATION 804
Function description
SUPERVISION → SYSTEM → OPERATION
CURRENT SYSTEM
CONDITION
(8040)
Use this function to check the current system condition.
Note!
All error messages are described in detail in the Operating Instructions for
PROline Prosonic Flow 93 FOUNDATION Fieldbus, BA 078D/06/en.
Display:
“SYSTEM OK” or the fault/notice message with the highest priority.
PREVIOUS SYSTEM
CONDITIONS
(8041)
Use this function to view the fifteen most recent fault and notice messages
since measuring last started.
Display:
The last 15 fault/notice messages appear on the display.
SIMULATION
FAILSAFE MODE
(8042)
Use this function to set all totalizers to their defined failsafe modes, in order to
check whether they respond correctly.
During this time, the words “SIMULATION FAILSAFE MODE” appear on the
display.
Options:
ON
OFF
Factory setting:
OFF
Note!
Active simulation is relayed to downstream function blocks or higher-level
process control systems via the FOUNDATION Fieldbus by means of the
status “UNCERTAIN” for the output value OUT (AI Block).
SIMULATION
MEASURAND
(8043)
Use this function to activate the simulation of a measured variable.
The simulation value is determined in the VALUE SIMULATION MEASURAND
(8044) function. When simulation is active, the message “SIMULATION
MEASURAND” appears on the local display.
Options:
OFF
VOLUME FLOW CH1
SOUND VELOCITY CH1
Factory setting:
OFF
Caution!
• The measuring device cannot be used for measuring while this simulation is
in progress.
• The setting is not saved if the power supply fails.
Note!
• Active simulation is relayed to downstream function blocks or higher-level
process control systems via the FOUNDATION Fieldbus by means of the
status “UNCERTAIN” for the output value OUT (AI Block).
• The simulation is independent of the position of the jumper for the
simulation mode on the I/O board.