7 Block BASIC FUNCTION PROline Prosonic Flow 93 FOUNDATION Fieldbus
60 Endress+Hauser
7.2.4 Function group: LIQUID DATA
BASIC FUNCTION G
Þ
FOUNDATION FIELDBUS GGA
ß
PROCESS PARAM. CH1 GIA
Þ
CONFIGURATION 640
ß
ADJUSTMENT 648
ß
PIPE DATA 652
ß
LIQUID DATA 654
ß
PROCESS PARAM. CH2 GIB
Þ
CONFIGURATION 640
ß
ADJUSTMENT 648
ß
PIPE DATA 652
ß
LIQUID DATA 654
Function description
BASIC FUNCTION PROCESS PARAMETER (CH1...CH2) LIQUID DATA
LIQUID
(6540)
Use this function to select the liquid in the pipe.
Options:
WATER
SEA WATER
DISTILLED WATER
AMMONIA
ALCOHOL
BENZENE
BROMIDE
ETHANOL
GLYCOL
KEROSENE
MILK
METHANOL
TOLUOL
LUBE OIL
FUEL OIL
PETROL
OTHERS
Factory setting:
WATER
TEMPERATURE
(6541)
Use this function to enter the temperature of the liquid.
User input:
Fixed-point number -273.15...726.85 °C (0...1000 K)
Factory setting:
20 °C

Table of Contents

Other manuals for Endress+Hauser Proline Prosonic Flow 93

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Proline Prosonic Flow 93 and is the answer not in the manual?

Endress+Hauser Proline Prosonic Flow 93 Specifications

General IconGeneral
BrandEndress+Hauser
ModelProline Prosonic Flow 93
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals