PROline Prosonic Flow 93 FOUNDATION Fieldbus 4 Block QUICK SETUP
Endress+Hauser 19
4Block QUICK SETUP
Block Group
Function
groups
Functions
QUICK SETUP
(B)
ÞÞ
SETUP
SENSOR
(1001) P. 19
Þ
QUICK SETUP
COMMISSION.
(1002) P. 19
T-DAT SAVE/
LOAD
(1009) P. 19
Function description
QUICK SETUP
SETUP SENSOR
(1001)
Use this function to start the Quick Setup menu for mounting the ultrasonic
sensors.
Options:
YES
NO
Factory setting:
NO
Note!
You will find a flowchart of the Quick Setup SENSOR menu on Page 20. Further
information on Quick Setups is provided in the related Operating Instructions
for Prosonic Flow 93 FOUNDATION Fieldbus, (BA078D/06/en/).
QUICK SETUP
COMMISSIONING
(1002)
Use this function to start the Quick Setup for commissioning.
Options:
YES
NO
Factory setting:
NO
Note!
You will find a flowchart of the COMMISSIONING Quick Setup menu on
Page 22. Further information on Quick Setups is provided in the related
Operating Instructions for Prosonic Flow 93 FOUNDATION Fieldbus,
(BA078D/06/en/).
T-DAT SAV E/ LOA D
(1009)
Use this function to save the parameter settings / configuration of the
transmitter in a transmitter DAT (T-DAT), or to load the parameter settings from
the T-DAT into the EEPROM (manual security function).
Application examples:
• After commissioning, the current measuring point parameters can be saved
to the T-DAT as a backup.
• If the transmitter is replaced for some reason, the data from the T-DAT can
be loaded into the new transmitter (EEPROM).
Options:
CANCEL
SAVE (from EEPROM to T-DAT)
LOAD (from the T-DAT into EEPROM)
Factory setting:
CANCEL
Note!
If the power supply fails, the totalizer readings are automatically saved to the
EEPROM.