PROline Prosonic Flow 93 FOUNDATION Fieldbus 8 Block SUPERVISION
Endress+Hauser 73
8.2 Group SYSTEM CH2
8.2.1 Function group: OPERATION
SUPERVISION J
Þ
SYSTEM JAA
ß
SYSTEM CH2 JAB
Þ
OPERATION 804
Function description
SUPERVISION → SYSTEM CH2 → OPERATION
SIMULATION
MEASURAND
(8043)
Use this function to activate the simulation of a measured variable.
The simulation value is determined in the VALUE SIMULATION MEASURAND
(8044) function. When simulation is active, the message “SIMULATION
MEASURAND” appears on the local display.
Options:
OFF
VOLUME FLOW CH2
SOUND VELOCITY CH2
Factory setting:
OFF
Caution!
• The measuring device cannot be used for measuring while this simulation is
in progress.
• The setting is not saved if the power supply fails.
Note!
• Active simulation is relayed to downstream function blocks or higher-level
process control systems via the FOUNDATION Fieldbus by means of the
status “UNCERTAIN” for the output value OUT (AI Block).
• The simulation is independent of the position of the jumper for the
simulation mode on the I/O board.
VALUE SIMULATION
MEASURAND
(8044)
Note!
This function is not displayed unless the SIMULATION
MEASURAND (8043) function is active.
Use this function to specify a selectable value, (e.g. 12 m³/s). This is used to
test the associated functions in the flowmeter itself and downstream
FOUNDATION Fieldbus function blocks.
User input:
5-digit floating-point number, [unit]
Factory setting:
0 [unit]
Caution!
• The setting is not saved if the power supply fails.
• The appropriate unit is taken from the SYSTEM UNITS function group, (ACA,
see Page 15).