5 Discrete Output function block PROline Prosonic Flow 93 FOUNDATION Fieldbus
186 Endress+Hauser
FSTATE_TIME AUTO - MAN -
OOS - CAS -
RCAS
Entry of the delay time which must expire with a fault of the
external reference value (RCAS_IN_D parameter), before the
defined fault operation is initiated.
Factory setting:
0 s
Note!
After expiry of the delay time and a still active fault, the following
values are used for further processing:
• If the option "Fault State to value" was selected in the IO_OPTS
parameter, the value of the FSTATE_VAL_D (fault replacement
value) parameter will be used.
• If the option "Fault State to value" was not selected in the
IO_OPTS parameter, the value of the RCAS_IN_D parameter
will continue to be used.
FSTATE_
VAL_D
AUTO - MAN -
OOS - CAS -
RCAS
Entry of a fault condition replacement value for the RCAS_IN_D
parameter.
This value is used for further processing instead of the
RCAS_IN_D parameter when the following conditions are met::
• An active fault of the RCAS_IN_D parameter exists.
• The delay time FSTATE_TIME has elapsed.
• The "Fault state to value" option has been selected in the
IO_OPTS parameter.
Options:
0 = Starting point, no effect
1 = Reserved
2 = Positive zero return channel 1 "ON"
3 = Positive zero return channel 1 "OFF"
4 = Zero point adjustment channel 1
5 = Reserved
6 = Reserved
7 = Reset totalizers 1, 2, 3
8 = Reset totalizer 1
9 = Reset totalizer 2
10 = Reset totalizer 3
11 = Reserved
12 = Reserved
13 = Reserved
14 = Reserved
15 = Reserved
16 = Positive zero return channel 2 "ON"
17 = Positive zero return channel 2 "OFF"
18 = Zero point adjustment channel 2
Factory setting:
0
GRANT_
DENY
AUTO - MAN -
OOS - CAS -
RCAS
Releases or restricts the access authorisation of a fieldbus host
system to the field device.
Note!
This parameter is not evaluated by PROline Prosonic Flow 93
FOUNDATION Fieldbus.
Discrete Output function block / base index 780
Parameter Write access
with operating
mode
(MODE_BLK)
Description