J22 TDLAS Gas Analyzer 4900002277
SpectraSensors, Inc. 119
110 Peak track adjust-
ment exceeded
1. Check process.
2. Check spectrum.
3. Reset peak tracking.
FAlarm
111 Peak track adjust-
ment warning
1. Check process.
2. Check spectrum.
3. Reset peak tracking.
FAlarm
Diagnostic of electronic
201 Device failure
1. Restart device.
2. Contact service.
FAlarm
242 Software incompatible
1. Check software.
2. Flash or change main electronics
module.
FAlarm
252 Modules incompatible
1. Check electronic modules.
2. Change electronic modules.
FAlarm
262 Sensor electronic
connection faulty
1. Check or replace connection cable
between sensor electronic module
(ISEM) and main electronics.
2. Check or replace ISEM or main elec-
tronics.
FAlarm
270 Main electronic failure Change main electronic module. F Alarm
271 Main electronic failure
1. Restart device.
2. Change main electronic module.
FAlarm
272 Main electronic failure
1. Restart device.
2. Contact service.
FAlarm
273 Main electronic failure Change electronic. F Alarm
275 I/O module 1 to n
defective
Change I/O module. F Alarm
276 I/O module 1 to n
faulty
1. Restart device.
2. Change I/O module.
FAlarm
283 Memory content
1. Reset device.
2. Contact service.
FAlarm
300 Sensor electronics
(ISEM) faulty
1. Restart device.
2. Replace sensor electronics.
FAlarm
Diagnostic
Number
Short Text Remedy Instructions
Status Signal
(from factory)
Diagnostic
Behavior
(from factory)