J22 TDLAS Gas Analyzer 4900002277
SpectraSensors, Inc. 73
Parameter overview with brief description
Carrying out additional display configurations
In the Display submenu you can set all the parameters associated with the
configuration of the local display.
Navigation
"Setup" menu → Advanced setup → Display
Parameter Description User Entry Factory Setting
SCC Enables or disables the Stream
Change Compensation feature.
•On
•Off
Off
Methane CH4 Sets the mole fraction of Methane
in the dry gas mixture.
0.4 to 1.0 mol 0.75 mol
Ethane C2H6 Sets the mole fraction of Ethane in
the dry gas mixture.
0.0 to 0.2 mol 0.1 mol
Propane C3H8 Sets the mole fraction of Propane
in the dry gas mixture.
0.0 to 0.15 mol 0.05 mol
IButane C4H10 Sets the mole fraction of Ibutane in
the dry gas mixture.
0.0 to 0.1 mol 0 mol
N-Butane C4H10 Sets the mole fraction of N-Butane
in the dry gas mixture.
0.0 to 0.1 mol 0 mol
Isopentane C5H12 Sets the mole fraction of Isopen-
tane in the dry gas mixture.
0.0 to 0.1 mol 0 mol
N-Pentane C5H12 Sets the mole fraction of N-Pen-
tane in the dry gas mixture
0.0 to 0.1 mol 0 mol
Neopentane C5H12 Sets the mole fraction of Neopen-
tane in the dry gas mixture
0.0 to 0.1 mol 0 mol
Hexane+ C6H14+ Sets the mole fraction of Hexane+
in the dry gas mixture
0.0 to 0.1 mol 0 mol
Nitrogen N2 Sets the mole fraction of Nitrogen
in the dry gas mixture.
0.0 to 0.55 mol 0 mol
Carbon diox. CO2 Sets the mole fraction of Carbon
dioxide in the dry gas mixture.
0.0 to 0.3 mol 0.1 mol
Hydrog.sulf. H2S Sets the mole fraction of Hydrogen
sulfide in the dry gas mixture.
0.0 to 0.05 mol 0 mol