J22 TDLAS Gas Analyzer 4900002277
SpectraSensors, Inc. 57
Setting peak tracking
The peak tracking submenu controls the software utility to keep the absorption
peak at a known location.
Navigation
"Setup" menu Peak Tracking
Parameter overview with brief description
Configuring the communication interface
The Communication submenu guides you systematically through all the
parameters that have to be configured for selecting and setting the
communication interface.
Analyzer control
Dew point
Reset
Average number
Parameter Prerequisite Description User Entry Factory Setting
Analyzer control Sets peak tracking
capability to OFF or
ON.
•Off
•On
Off
Reset Used if peak track-
ing is set above.
Resets peak tracking. Off
Reset
Off
Average number Used if peak track-
ing is set above.
Sets number of mea-
surements before
making a peak track-
ing adjustment.
Positive inte-
ger
10

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser SpectraSensors J22 TDLAS and is the answer not in the manual?

Endress+Hauser SpectraSensors J22 TDLAS Specifications

General IconGeneral
BrandEndress+Hauser
ModelSpectraSensors J22 TDLAS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals