4900002277 J22 TDLAS Gas Analyzer
56 SpectraSensors, Inc.
Setting the dew point
The dew point submenu configures parameters needed to perform a moisture dew
point calculation.
Navigation
"Setup" menu → Dew point
Parameter overview with brief description
Date/time format Sets the display unit for Date/time
format.
Unit Choose list • dd.mm.yy
hh:mm
• mm/dd/yy
hh:mm am/pm
Parameter Description User Entry Factory Setting
Dew pt method 1
Dew point
Dew pt method 2
Conversion type
Pipeline press.
Parameter Prerequisite Description User Entry Factory Setting
Dew pt method 1 — Sets the method used to cal-
culate dew point tempera-
ture.
•ASTM1
•ASTM2
•AB
ASTM2
Dew pt method 2 — Sets the method used to cal-
culate dew point tempera-
ture.
•ASTM1
•ASTM2
•AB
•
Off
Conversion type Used if Dew pt is
enabled by select-
ing a method
above.
Sets conversion type used to
calculate dew point tem-
perature.
• Ideal
•Real
Ideal
Pipeline press. Used if Dew pt is
enabled by select-
ing a method
above.
Sets the pressure at which
the dew point temperature
is calculated.
Floating point
number
• 50000 mbar a
• 725 psi a