EasyManuals Logo

Endress+Hauser SpectraSensors J22 TDLAS User Manual

Endress+Hauser SpectraSensors J22 TDLAS
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
4900002277 J22 TDLAS Gas Analyzer
56 SpectraSensors, Inc.
Setting the dew point
The dew point submenu configures parameters needed to perform a moisture dew
point calculation.
Navigation
"Setup" menu Dew point
Parameter overview with brief description
Date/time format Sets the display unit for Date/time
format.
Unit Choose list dd.mm.yy
hh:mm
mm/dd/yy
hh:mm am/pm
Parameter Description User Entry Factory Setting
Dew pt method 1
Dew point
Dew pt method 2
Conversion type
Pipeline press.
Parameter Prerequisite Description User Entry Factory Setting
Dew pt method 1 Sets the method used to cal-
culate dew point tempera-
ture.
•ASTM1
•ASTM2
•AB
ASTM2
Dew pt method 2 Sets the method used to cal-
culate dew point tempera-
ture.
•ASTM1
•ASTM2
•AB
Off
Conversion type Used if Dew pt is
enabled by select-
ing a method
above.
Sets conversion type used to
calculate dew point tem-
perature.
Ideal
•Real
Ideal
Pipeline press. Used if Dew pt is
enabled by select-
ing a method
above.
Sets the pressure at which
the dew point temperature
is calculated.
Floating point
number
50000 mbar a
725 psi a

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser SpectraSensors J22 TDLAS and is the answer not in the manual?

Endress+Hauser SpectraSensors J22 TDLAS Specifications

General IconGeneral
BrandEndress+Hauser
ModelSpectraSensors J22 TDLAS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals