J22 TDLAS Gas Analyzer 4900002277
SpectraSensors, Inc. 143
Local Operation Via display module
A0026785
1: Operation with touch control
Display elements
• 4-line, illuminated, graphic display
• White background lighting; switches to red in event of device errors
• Format for displaying measured variables and status variables can be individ-
ually configured
• Permitted ambient temperature for the display: –20 to +60 °C (–4 to +140 °F)
The readability of the display may be impaired at temperatures outside the
temperature range.
Operating elements
• External operation via touch control (3 optical keys) without opening the
housing: , ,
• Operating elements also accessible in various hazardous areas
J22 TDLAS Gas Ana-
lyzer with Enclosed
SCS
1
, with Heater
cCSAus
: Ex db ia op is IIC T3 Gb
Class I, Zone 1, AEx db ia op is IIC T3 Gb
Class I, Division 1, Groups B, C, D, T3
Tambient = -20 C to +60 C
ATEX/IECEx
: II 2G
Ex db ib op is h IIC T3 Gb
Tambient = -20 C to +60 C
Ingress Protection IP66, Type 4X
1 Sample Conditioning System