J22 TDLAS Gas Analyzer 4900002277
SpectraSensors, Inc. 121
387 HistoROM data faulty Contact service organization. F Alarm
410 Data transfer
1. Check connection.
2. Retry data transfer.
FAlarm
412 Processing download Download active, please wait. C Warning
431 Trim 1 to n Carry out trim. C Warning
437 Configuration incom-
patible
1. Restart device.
2. Contact service.
FAlarm
438 Dataset
1. Check data set file.
2. Check device configuration.
3. Up- and download new configuration.
M Warning
441 Current output 1 to n
1. Check process.
2. Check current output settings.
S Warning
484 Failure mode
simulation
Deactivate simulation. C Alarm
491 Current output 1 to n
simulation
Deactivate simulation. C Warning
494 Switch output simula-
tion 1 to n
Deactivate simulation switch output. C Warning
495 Diagnostic event simu-
lation
Deactivate simulation. C Warning
500 Laser current out of
range
1. Check spectrum.
2. Reset peak tracking.
M, C Warning
501 SCC configuration
faulty
1. Check gas composition settings.
2. Check gas composition sum.
C Warning
520 I/O 1 to n hardware
configuration invalid
1. Check I/O hardware configuration.
2. Replace wrong I/O module.
3. Plug the module of double pulse out-
put on correct slot.
FAlarm
537 Configuration
1. Check IP addresses in network.
2. Change IP address.
F Warning
594 Relay output simula-
tion
Deactivate simulation switch output. C Warning
Diagnostic
Number
Short Text Remedy Instructions
Status Signal
(from factory)
Diagnostic
Behavior
(from factory)