4900002277 J22 TDLAS Gas Analyzer
76 SpectraSensors, Inc.
100% bargraph
value 3
A selection was
made in the Value 3
display parameter.
Enter 100% value for
bar graph display.
Signed floating-
point number
1700 mbar a
Decimal places 3 A measured value
is specified in the
Value 3 display
parameter.
Select the number of
decimal places for the
display value.
• x
•x.x
•x.xx
•x.xxx
•x.xxxx
x.xx
Value 4 display A local display is
provided.
Select the measured
value that is shown on
the local display.
For the picklist,
see the Value 2
display parameter
Cell gas temperature
Decimal places 4 A measured value
is specified in the
Value 4 display
parameter.
Select the number of
decimal places for the
display value.
• x
•x.x
•x.xx
•x.xxx
•x.xxxx
x.xx
Display language A local display is
provided.
Set display language English English
Display interval A local display is
provided.
Set time measured val-
ues are shown on dis-
play if display
alternates between
values.
1 to 10 s 5 s
Display damping A local display is
provided.
Set display reaction
time to fluctuations in
the measured value.
0.0 to 999.9 s 0.0 s
Header A local display is
provided.
Select header contents
on local display.
• Device tag
• Free text
Device tag
Header text In the Header
parameter, the
Free text option is
selected.
Enter display header
text.
Max. 12 charac-
ters such as let-
ters, numbers or
special characters
(e.g. @, %, /)
------------
Separator A local display is
provided.
Select decimal separa-
tor for displaying
numerical values.
•. (point)
• , (comma)
. (point)
Parameter Prerequisite Description User Entry Factory Setting