J22 TDLAS Gas Analyzer 4900002277
SpectraSensors, Inc. 81
Navigation
"Diagnostics" menu → Simulation
Parameter overview with brief description
Simulation
Value current input 1 to n
Current output 1 to n simulation
Value current output 1 to n
Switch output simulation 1 to n
Switch status 1 to n
Relay output 1 to n simulation
Switch status 1 to n
Device alarm simulation
Diagnostic event category
Diagnostic event simulation
Parameter Prerequisite Description User Entry Factory Setting
Current output 1
to n simulation
— Switch the simulation
of the current output
on and off.
•Off
•On
Off
Value current out-
put 1 to n
In the Current out-
put 1 to n simula-
tion parameter, the
On option is
selected.
Enter the current value
for simulation.
3.59 to 22.5 mA 3.59 mA
Switch output sim-
ulation 1 to n
In the Operating
mode parameter,
the Switch option is
selected.
Switch the simulation
of the switch output on
and off.
•Off
•On
Off