J22 TDLAS Gas Analyzer 4900002277
SpectraSensors, Inc. 89
Navigation
"Diagnostics" menu Data logging
Parameter overview with brief description
Logging interval
Clear data logging
Data logging
Logging delay
Data logging control
Data logging status
Entire logging duration
Assign channel 1...4
Data logging
Parameter Prerequisite Description
Selection/ User
Interface/User
Entry
Factory Setting
Assign channel 1
to n
The Extended His-
toROM application
package is avail-
able.
Assign process vari-
able to logging chan-
nel.
•Off
Concentration
1)
•Dew point 1
•Dew point 2
•Cell gas
pressure
•Cell gas
temperature
•Current output
1 to n
Off
Logging interval The Extended
HistoROM applica-
tion package is
available.
Define the logging
interval for data log-
ging. This value
defines the time
interval between the
individual data points
in the memory.
0.1 to 999.0 s 1.0 s