18.3.2 Correction of the flow velocity
After the profile bounds have been defined, it is possible to define a correction of the flow
velocity:
v
cor
= m · v + n
where
All quantities derived from the flow velocity will be calculated with the corrected flow ve-
locity. The correction data will be transmitted to the PC or printer during the online or
offline transmission.
Select on to define the correction data, off to work without correction of the flow veloc-
ity and return to the menu item SYSTEM settings.
•If on is selected, enter the slope. The input 0.0 deactivates the correction.
Press ENTER.
Enter the offset. Enter zero to work without offset.
Press ENTER.
v measured flow velocity
m slope, range: -2.0...+2.0
n offset, range: -12.7…+12.7 cm/s
v
cor
corrected flow velocity
Notice!
It will not be displayed that the correction of the flow velocity is active during the
measurement.
Special Funct.\...\Calibration\...\Calibration
Special Funct.\...\Calibration\...\Calibration\Slope
Special Funct.\...\Calibration\...\Calibration\Offset

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F60 Series and is the answer not in the manual?

Flexim FLUXUS F60 Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F60 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals