9 Measurement
9.1 Parameter input FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
86
9.1.2 Input of fluid parameters
Fluid
Select the fluid from the scroll list.
Press ENTER.
If the fluid is not in the scroll list, select Other Medium.
It is possible to specify which fluids will be displayed in the scroll list, see section 19.4.
If a fluid is selected from the scroll list, the menu item for the input of the fluid temperature
is displayed directly.
If Other Medium is selected, the fluid parameters have to be entered first:
average sound speed of the fluid
range around the average sound speed of the fluid
kinematic viscosity
density
Sound speed of the fluid
The sound speed of the fluid is used for the calculation of the transducer distance. The
exact value of the sound speed is not always known. Therefore, a range of possible val-
ues for the sound speed has to be entered.
Enter the average sound speed of the fluid.
Press ENTER.
This display will only be indicated if Other Medium is selected.
Sound speed range of the fluid
Select auto if the area around the average sound speed is to be calculated by the
transmitter.
Select user if the range around the average sound speed has to be entered.
Press ENTER.
This display will only be indicated if Other Medium is selected.
Parameter\Medium
Parameter\Medium\Other Medium\c-Medium
Parameter\Medium\Other Medium\c-Medium range

Table of Contents