9 Measurement
9.2 Measurement settings FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
90
For the selected physical quantity (except for the sound speed), a scroll list with the avail-
able units of measurement is displayed. The unit of measurement which was selected
previously is displayed first.
Select the unit of measurement of the selected physical quantity.
Press ENTER.
9.2.2 Input of the damping factor
Each displayed measured value is a floating average of all measured values of the last x
seconds, with x being the damping factor. A damping factor of 1 s means that the mea-
sured values are not averaged because the measuring rate is approx 1/s. The default
value of 10 s is appropriate for normal flow conditions. Values which fluctuate strongly
due to a higher flow dynamic, require a higher damping factor.
Select the program branch Output Options.
Press ENTER until the menu item Damping is displayed.
Enter the damping factor.
Press ENTER.
Press key BRK to return to the main menu. The further displays of the program branch
Output Options are used for the activation of the measured value transmission.
Output Options\for Channel A:\Physic. Quant.\Volume flow
Notice!
If the physical quantity or the unit of measurement is changed, the settings of the
outputs have to be checked, see section 9.2.3.
Output Options\for Channel A:\...\Damping

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F60 Series and is the answer not in the manual?

Flexim FLUXUS F60 Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F60 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals