EasyManua.ls Logo

GE AT600 - Page 109

GE AT600
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AT600 User’s Manual 97
Chapter 5. Communication
Command 153 (0x99): Write Density Value
This command will write density value in meter.
Table 36: Request Data Bytes for Write Density Value
Byte Format Description
0 Unsigned-8 Density type:
1: Actual Density;
2: Reference Density;
1 Unsigned-8 Density Unit Code
2 - 5 Float Density value
Table 37: Response Data Bytes for Write Density Value
Byte Format Description
0 Unsigned-8 Density type:
1: Actual Density;
2: Reference Density;
1 Unsigned-8 Density Unit Code
2 - 5 Float Density value
Table 38: Command-Specific Response Codes for Write Density Value
Code Class Description
0 Success No Command-Specific Errors
1Undefined
2 Error Invalid Selection
3 - 4 Undefined
5 Error Too Few Data Bytes Received
6 Error Device-Specific Command Error
7Error In Write Protect Mode
8 - 15 Undefined
16 Error Access Restricted
17 - 127 Undefined

Table of Contents