Chapter 5. Communication
140 AT600 User’s Manual
Command 232 (0xE8): Write Error Limits
This command is to write flow meter error limits.
Table 165: Request Data Bytes for Write Error Limits
Byte Format Description
0 Unsigned-8 Error limit :
Correlation Peak Limit
Acceleration Limit
Velocity Low Limit
Velocity High Limit
Amp Disc Min
Amp Disc Max
Signal Low Limit
Sound Speed Limit
Errors Allowed
1 - 4 float Error limit Value;
Table 166: Response Data Bytes for Write Error Limits
Byte Format Description
0 Unsigned-8 Error limit:
Correlation Peak Limit
Acceleration Limit
Velocity Low Limit
Velocity High Limit
Amp Disc Min
Amp Disc Max
Signal Low Limit
Sound Speed Limit
Errors Allowed
1 - 4 float Error limit Value;