AT600 User’s Manual 141
Chapter 5. Communication
Command 232 (0xE8): Write Error Limits (cont.)
Command 233 (0xE9): Write Signal Setup
This command is to write flow meter signal setup.
Table 167: Command-Specific Response Codes for Write Error Limits
Code Class Description
0 Success No Command-Specific Errors
1Undefined
2 Error Invalid Selection
3-4 Undefined
5 Error Too Few Data Bytes Received
6 Error Device-Specific Command Error
7Error In Write Protect Mode
8-15 Undefined
16 Error Access Restricted
17-127 Undefined
Table 168: Request Data Bytes for Write Signal Setup
Byte Format Description
0 Unsigned-8 signal setup type:
Delta T Offset
percentage Peak
Min Peak Percentage
Max Peak percentage
1 - 4 Float signal setup Value
Table 169: Response Data Bytes for Write Signal Setup
Byte Format Description
0 Unsigned-8 signal setup type:
Delta T Offset
percentage Peak
Min Peak Percentage
Max Peak percentage