AT600 User’s Manual 143
Chapter 5. Communication
Command 239 (0xEF): Reset Flow Meter Data (cont.)
Command 241 (0xF1): Read the Factory Setting
This command is to read the factory setting.
Table 172: Response Data Bytes for Reset Flow Meter Data
Byte Format Description
0 Unsigned-8 Reset type:
Reset Error Log
Forward Inventory
Reverse Inventory
Net Inventory
Inventory Time
All
Inventory
Table 173: Command-Specific Response Codes for Reset Flow Meter Data
Code Class Description
0 Success No Command-Specific Errors
1Undefined
2 Error Invalid Selection
3-4 Undefined
5 Error Too Few Data Bytes Received
6 Error Device-Specific Command Error
7Error In Write Protect Mode
8-15 Undefined
16 Error Access Restricted
17-127 Undefined
Table 174: Request Data Bytes for Read the Factory Setting
Byte Format Description
None