Chapter 3. Initial Setup and Programming
46 AT600 User’s Manual
Setting Pulse Error Handling
Test Pulse
To change the
Error Handling status, select it on the screen and press []. The
following screen appears.
Select
Hold or Stop. Hold directs the meter, in case of a flow measurement error,
to keep sending the pulses sent at the last good reading. Stop directs the meter, in
case of measurement error, to stop pulsing.
Press
[], and the screen returns to the previous display. Press [] to return to the
Digital Output menu.
To test the pulse output, select
Test Pulse and press []. The following screen
appears.
Press
[]. A screen similar to the following appears.
Press
[], and the screen returns to the previous display. Press [] to return to the
Digital Output menu.
Enter a number of pulses and the instrument will send that many out. Note on
your measurement system that the right number of pulses were received.
After testing,
press [] to return to the Digital Output menu.
Pulse Value
Pulse Time
Test Pulse
Pulse Value
Pulse Time
Test Pulse
Test Pulse