Chapter A5: Image Viewer
Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) A5-3
Operator Manual © 2008 General Electric Company. All rights reserved.
tool selection list Switches between different tool panels that allow you to change
the image display or manipulate the image.
Available options are:
• Images – Processed: Shows small previews of all processed
images in the selected series and highlights the images
currently shown in the Viewer. The images are categorized by
exam and series. Refer to Select Images to View
(p. A5-5).
• Images – Raw: Shows small previews of all raw images in the
selected series and highlights the images currently shown in
the viewer. The images are categorized by exam and series.
Refer to Select Images to View
(p. A5-5).
• Format/Zoom: Controls how many images appear in the
Image Viewer screen at one time and adjusts the
magnification of each image. Refer to Change Image Display
(p. A5-6).
• Image Display Tools: Contains the tools to flip, rotate, adjust
brightness, adjust contrast, invert, and apply windowing to
images. Refer to Adjust Images
(p. A5-7).
• Annotations/Mask: Contains the tools to annotate and crop
images. Refer to Annotate and Mask Images
(p. A5-9).
• Reprocessing: Changes the processing settings. Refer to Re-
process Images (p. A5-17).
• Print: Controls the printing of images. Refer to Print Images
(p. A5-22).
DEI (Detector Exposure Index) If enabled, displays the dose received by the detector and
whether the dose is within an acceptable range for the
anatomy.
You may need to re-take images that show doses below the
acceptable range. Refer to Determine if Images should be
Retaken (DEI) (p. A5-18) for more information.
NOTE: Depending on your system’s configuration, the DEI may
only show a numerical value. Refer to Chapter A7: Set
Preferences-Image Viewer (p. A7-25) to configure the
Detector Exposure Index.
pointer controls Changes the action of the pointer when pressed and dragged
on the image.
Refer to Change Pointer Controls
(p. A5-21) for more
information.
Function Description